Product Features
• Oxidation method (High efficiency oxidation reactor considering UV transmission distance, UV light amount detection and replacement alarm function)
• Measurement conditions can be varied (change conditions such as IC removal time, Acid injection amount, UV lamp, carrier gas injection amount, etc.)
• It has a long UV lamp life, making it easy to maintain.
• Suitable for analysis of samples containing high turbidity and suspended matter
• Widely applicable from river water to sewage and wastewater
Product Specifications
Measurement range | 0~100 mg/L (Sample Dilution : 1,000/10,000 mg/L) | Electrical Output Signal | 4-20 mA, RS-232C, TCP/IP |
Minimum unit of scale | 0.001 mg/L | Power Supply | AC 220V, 50-60Hz, Power consumption: 100VA |
Measurement method | UV wet oxidation method (NPOC method) | Temperature Range / Humidity | 5~40 °C, 85 % RH or less |
Repeatability / Accuracy | ±2% of F.S | Material | Steel, Powder Coating |
Oxidation method | Persulfate + UV + Thermal Oxidation | Dimensions | 550(W) x 430(D) x 1,560(H )mm |
Sample / Reagent measurement | Sample loop Line measuring | Weight | 80 kg |
Channel | Up to 6 channels (optional) | Operating Temperature | 5~40 °C |
Measurement time | Within 10 minutes | Alarm List | No sample, No distilled water, No reagent, No carrier gas, UV-Lamp abnormal, MFC abnormal, Detector abnormal, System abnormal, Consumption replacement |
Display method | Color LCD (10.2") Touch Screen | ||
Data Storage | Store PC HDD, USB memory stick available | ||
Status Output | Normal, Calibrating, Operating error, Temporary pause | Carrier Gas | High purity nitrogen gas or Air, Supply Pressure: 0.1~0.15 Mpa |
Reagent Replacement Cycle | 4 weeks (2L) | Option | Carrier gas generator |
Core Technology for TOC 2200
Classification | Before | After |
Sample Amount | 15 ml | 5 ml |
Wash Amount | 345 ml | 20 ml |
Reagent Method | Mechanical Flow Controller | Electronic MFC |
Gas Regulator | 7 days | 30 days |
Controller | Manual | Automatic |
Repeatability | 2.60 % | 1.24 % |