본문 바로가기 주메뉴 바로가기

Products

Accurate water quality, reliable partner.

Water Quality TMS Monitor

TOC 2200
TOC 2200
TOC 2200
TOC 2200

TOC 2200

Product Features

Oxidation method (High efficiency oxidation reactor considering UV transmission distance, UV light amount detection and replacement alarm function)

Measurement conditions can be varied (change conditions such as IC removal time, Acid injection amount, UV lamp, carrier gas injection amount, etc.)

It has a long UV lamp life, making it easy to maintain.

Suitable for analysis of samples containing high turbidity and suspended matter

Widely applicable from river water to sewage and wastewater


Product Information
Key Features
DISPLAY PART
• Check the operation status of the measuring instrument
• Measuring instrument CONTROL PART

Measurement mechanism
• Optimum oxidation conditions
• Oxidant injection
• UV light source contact
• Oxidator sample contact

Reagent Storage
• Analytical Reagents
• STD SOLUTION

Detailed Product Description

Product Specifications

Measurement range

0~100 mg/L

(Sample Dilution : 1,000/10,000 mg/L)

Electrical Output Signal
4-20 mA, RS-232C, TCP/IP
Minimum unit of scale
0.001 mg/L
Power Supply
AC 220V, 50-60Hz, Power consumption: 100VA
Measurement method

UV wet oxidation method (NPOC method)

Temperature Range / Humidity
5~40 °C, 85 % RH or less
Repeatability / Accuracy
±2% of F.S
Material
Steel, Powder Coating
Oxidation method
Persulfate + UV + Thermal Oxidation
Dimensions
550(W) x 430(D) x 1,560(H )mm
Sample / Reagent measurement
Sample loop Line measuring
Weight
80 kg
Channel
Up to 6 channels (optional)
Operating Temperature
5~40 °C
Measurement time
Within 10 minutes
Alarm List
No sample, No distilled water, No reagent, No carrier gas, UV-Lamp abnormal, MFC abnormal, Detector abnormal, 
System abnormal, Consumption replacement
Display method
Color LCD (10.2") Touch Screen
Data Storage
Store PC HDD, USB memory stick available
Status Output
Normal, Calibrating, Operating error, Temporary pause
Carrier Gas
High purity nitrogen gas or Air, 
Supply Pressure: 0.1~0.15 Mpa
Reagent Replacement Cycle
4 weeks (2L)
Option
Carrier gas generator



Core Technology for TOC 2200

Classification
Before
After
Sample Amount
15 ml
5 ml
Wash Amount
345 ml
20 ml
Reagent Method
Mechanical Flow Controller
Electronic MFC
Gas Regulator
7 days
30 days
Controller
Manual
Automatic
Repeatability
2.60 %
1.24 %